i-STAT ACT Kaolin Cartridges in vitro diagnostic test which uses fresh whole blood to monitor high-dose heparin anticoagulation frequently associated with cardiovascular surgery.
This recall has been terminated (originally issued November 17, 2010).
- Company
- Abbott Point Of Care Inc.
- Recall Initiated
- October 6, 2010
- Posted
- November 17, 2010
- Terminated
- August 22, 2012
- Recall Number
- Z-0400-2011
- Quantity
- 95,325 cartridges
- Firm Location
- Princeton, NJ
- Official Source
- View on FDA website ↗
Reason for Recall
Certain lots of i-STAT Celite ACT and Kaolin ACT cartridges may be difficult to fill or will not fill when attempting to transfer blood to the sample entry port of the cartridge.
Distribution
Worldwide Distribution
Lot / Code Info
List numbers: 07G81-01, 07G81-02; Lot numbers: S10129, S10129A, S10130, S10130B, S10147A, S10147B, S10147C, S10147D, S10147E, S10148A, S10148B, S10148C, S10148D, S10148G, S10149, S10149B, S10149C, S10151A, S10151B, S10152, S10152A, S10152B, S10152C, S10152D, S10152E, S10152F, S10154A, S10154B, S10154D, S10160B, S10162.
Root Cause
Other
Action Taken
Abbott Point of Care, Inc. sent an URGENT RECALL NOTICE dated September 2010 on October 6, 2010 by Federal Express to all US customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of cartridges from the identified lots. If customers did not currently have replacement cartridges then they should continue to use the cartridge lot until their facility receives replacement material. Upon receipt of replacement material, customers were to discontinue use of and return the affected cartridges and switch to the replacement material. All unused cartridges will be credited and will be returned. If you have any questions regarding this information, please contact Abbott Point of Care Technical Support at 800-366-8020, Option 1 or your Abbott Point of Care representative.