ARTISTE, ONCOR and PRIMUS Linear Accelerators with RT Therapist Workspace, version 4.1. Manufactured by Siemens AG, Kemnath, Germany. The indication is to deliver X-ray, photon and electron radiation for the therapeutic treatment of cancer.

This recall has been terminated (originally issued February 23, 2012).

Company
Siemens Medical Solutions Usa, Inc
Recall Initiated
January 22, 2012
Posted
February 23, 2012
Terminated
November 7, 2012
Recall Number
Z-1092-2012
Quantity
182 active devices
Firm Location
Concord, CA
Official Source
View on FDA website ↗

Reason for Recall

Software upgrade to installed base to the latest release of the RT Therapist version 4.1.144 which corrects multiple safety issues that could lead to patient mistreatment.

Distribution

Worldwide Distribution - USA, including the states of WI, UT, PA, TN, OH, LA, DE, MI, NJ, FL, NE, WV, MO and the countries of Germany, Spain, Poland, Australia, Malaysia, New Zealand, India, Republic of Korea, Canada, Italy, Egypt, Philippines, France, Belgium, Saudi Arabia, South Africa, Norway, Turkey , Hungary, Russian Federation, P.R. China, Lebanon, Iran, and the United Kingdom.

Lot / Code Info

Syngo RT Therapist v4.1, part number 8162815; Syngo RT Therapist Assist (off line system), part number 8162807; Syngo RT Express Basic (stand alone system), part number 8151289; and Syngo RT Express Assist (offline system), part number 8151297.

Root Cause

Software design

Action Taken

Siemens customers with active devices will receive the software update through the distribution of the "Update Instructions" with a release date of January 22, 2012. The software update will be delivered and installed by a qualified Field Service Engineer utilizing a SW kit and instruction distributed by the Update instruction. Consignee notification of the Update Instructions will be delivered by Customer Service or by certified mail.