RecallDepth

RapidPoint 400 Analyzers for Blood Gas, Electrolytes and metabolites Software Versions prior to 3.5.1. The Siemens 400 series system is intended for point of care and laboratory testing of blood gases, electrolytes and metabolites in arterial, venous and capillary whole blood samples.

Company
Siemens Healthcare Diagnostics Inc
Recall Initiated
October 30, 2009
Posted
March 8, 2010
Terminated
September 4, 2015
Recall Number
Z-1019-2010
Quantity
998 units
Firm Location
Norwood, MA

Reason for Recall

Software: Reference sensor failure may result in inaccurate pH reuslts

Distribution

Worldwide distribution: USA and Australia, Austria, Bahamas, Barbados, Belgium, Bosnia Herzeg, Brazil, Canada, Canary Islands, Chile, Croatia, Czech Republic, Egypt, France, Germany, Holy See (Vatican City State) Hong Kong, Ireland, Italy, Japan, Macedonia, Malaysia, M¿xico, New Zealand Not assigned P.R. China, Poland, Portugal, Puerto Rico, Refurb.Systems U.S.A Republic Korea, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, U.A.E., United Kingdom

Lot / Code Info

Software Versions prior to 3.5.1

Root Cause

Component design/selection

Action Taken

An "Urgent Field Safety Notice" dated September 2009 was sent via overnight mail to customers and included software Version 3.5.2 with instructions for install that will detect and prevent the reporting of potentially errant test results. The customers are requested to please install the included software update immediately and return the attached Fax back form. Siemens will provide the mandatory update to all customers with the Rapidpoint 400 /405 test system. If you have any questions, please contact your local Technical Support at 508-359-3877.

More recalls by Siemens Healthcare Diagnostics Inc

View all recalls by this company →