Proteus XR/a Radiographic X-ray System, models 2259973-1, 2259973-2, 2259973-3, 2259973-4. The system consists of a radiographic table, a overhead tube suspension, high frequency generator and power distribution unit, wall stand, dual focal spot x-ray tube and operator''s console.
This recall has been terminated (originally issued February 14, 2006).
- Company
- General Electric Med Systems Llc
- Recall Initiated
- December 30, 2004
- Posted
- February 14, 2006
- Terminated
- February 21, 2006
- Recall Number
- Z-0219-05
- Quantity
- 2716
- Firm Location
- Waukesha, WI
- Official Source
- View on FDA website ↗
Reason for Recall
Units have a defect in accordance with 21 CFR 1003.2 because an incorrect exposure can occur.
Distribution
globally
Lot / Code Info
all serial numbers
Root Cause
Other
Action Taken
An 'Urgent: Radiation Safety Notice' letter was issued to hospitals with affected systems describing potential hazards, the planned modification and recommended action to limit the hazard until the planned modification is installed. A GE field Service Representative will install the modification.