ACCU-CHEK Inform meter; catalog No. 3035123.
This recall has been terminated (originally issued July 20, 2004).
- Company
- Roche Diagnostics Corp.
- Recall Initiated
- June 30, 2004
- Posted
- July 20, 2004
- Terminated
- October 3, 2005
- Recall Number
- Z-1060-04
- Firm Location
- Indianapolis, IN
- Official Source
- View on FDA website ↗
Reason for Recall
Exposure to excessive disinfectant or cleaning solution may cause overheating, melting, or smoke.
Distribution
Nationwide.
Lot / Code Info
All units.
Root Cause
Other
Action Taken
A recall notice dated 6/30/04 was sent to each customer instructing them on proper cleaning methods, asking them to inspect their units for damage, and providing warning stickers for attachment to the base units.