GE Centricity PACS RA1000 Workstation; for diagnostic image analysis. GE Healthcare Integrated IT Solutions, Barrington, IL 60010

This recall has been terminated (originally issued September 2, 2008).

Company
Ge Healthcare Integrated It Solutions
Recall Initiated
May 6, 2008
Posted
September 2, 2008
Terminated
February 3, 2010
Recall Number
Z-1826-2008
Quantity
177 units
Firm Location
Barrington, IL
Official Source
View on FDA website ↗

Reason for Recall

A software anomaly results in incorrect study date and time information being displayed in the report screen and title, which may result in a potential patient misdiagnosis.

Distribution

Nationwide and internationally to Australia, Belgium, Canada, Denmark, France, Germany, Italy, Japan, Malta, South Korea, Spain, Switzerland and United Kingdom.

Lot / Code Info

GE Centricity PACS RA1000 Workstation software version: 3.0.0.X, 3.0.1.X, 3.0.2.X, and 3.0.3.X

Root Cause

Software design

Action Taken

GE Healthcare IITS sent Product Safety Notification letters dated 5/5/08 to all customers who have the GE Centricity PACS Software Versions 3.0.0.X, 3.0.1.X, 3.0.2.X and 3.0.3.X informing them of the incorrect study date and time information being displayed in the report screen and title, which may lead to a potential patient misdiagnosis. The users were instructed to refer to the image overlay for the correct date and time until the new software version can be installed. A software patch to permanently resolve the issue will be installed at no charge. Any questions were directed to the GE Customer Care Center at 800-437-1171.