Siemens Primus Linear Accelerator System, PRIMART MX Linear Accelerator System; PRIMUS Mid, Part Number:1940035 and PRIMART MX, Part Number: 5500371, Siemens Medical Solutions USA, Inc., Concord, CA
- Recall Initiated
- July 27, 2007
- Posted
- February 19, 2008
- Terminated
- April 29, 2011
- Recall Number
- Z-0850-2008
- Quantity
- 1,284 total for recalls Z-0850-0852-2008
- Firm Location
- Concord, CA
Reason for Recall
Door Hinge failures; loose or falling off (Stationary Structure Doors)
Distribution
Worldwide- USA including the states of Missouri, Georgia, Delaware, Texas, Ohio, Alabama, Kentucky, Florida, California, North Carolina, New Jersey, New York, Illinois, Pennsylvania, Colorado, Tennessee, Arizona, Pennsylvania, Virginia, New Jersey, Michigan, Wisconsin, Indiana, Nebraska, Montana, Kansas, Mississippi, Rhode Island, Washington, South Carolina, Maryland, New Hampshire, Maine, Arkansas, Idaho, Missouri, Louisiana, Oklahoma, North Dakota, Minnesota, Massachusetts, and The District of Columbia and countries of Canada and Mexico.
Lot / Code Info
All Serial Numbers up to Serial Number: 3532
Root Cause
Component design/selection
Action Taken
Stationary Door Hinges will be replaced on each device by the recalling firm's technician visits.