CLINITRON CII Air Fluidized Therapy Units. Intended to prevent and/or treat pressure ulcer development and wound deterioration in patients who have a significant risk of developing these problems and who generally also have one or more of the following conditions: immobility, poor nutrition, diminished level of consciousness, reduced subcutaneous tissue or multi-system failure.

Company
Hill-Rom Manufacturing, Inc.
Recall Initiated
June 24, 2009
Posted
August 25, 2009
Terminated
July 11, 2012
Recall Number
Z-1902-2009
Quantity
625 units (total both units)
Firm Location
Charleston, SC

Reason for Recall

Incorrectly repaired power cord wires can potentially overheat resulting in smoldering sound foam in the base of the therapy bed.

Distribution

Worldwide Distribution -- United States, Abu Dhabi, Brazil, Canada, Chile, Columbia, Mexico, Newfoundland, Puerto Rico, Saudi Arabia, Singapore, Taiwan, Turkey and Venezuela.

Lot / Code Info

All serial numbers.

Root Cause

Equipment maintenance

Action Taken

Hill-Rom issued an "Urgent Field Safety Notice" letter dated February 2, 2009 addressed to Facility Risk Managers and Facility Administrators. The notice detailed the issue and instructed users on actions to be taken. For further information, please contact Hill-Rom Customer Service at 1-800-638-2546.