Toshiba Digital Radiography System Model DFP8000D
- Company
- Toshiba American Med Sys Inc
- Recall Initiated
- November 15, 2005
- Posted
- February 23, 2006
- Terminated
- July 12, 2007
- Recall Number
- Z-0532-06
- Quantity
- 3
- Firm Location
- Tustin, CA
Reason for Recall
In certain conditions the FOV of the lateral plane does not coincide with the X-ray exposed area, and consequently a part of the X-ray field may not be displayed in the magnification mode.
Distribution
CO, MI, & TX
Lot / Code Info
Serial #s W1C0552175, W1C0552181, W1C0562189
Root Cause
Other
Action Taken
An Advisory Letter will be sent to customers via certified mail. Corrected software will be provided at no charge. Customers can contact their local Toshiba Service person to schedule the software update.