Toshiba Digital Radiography System Model DFP8000D

Company
Toshiba American Med Sys Inc
Recall Initiated
November 15, 2005
Posted
February 23, 2006
Terminated
July 12, 2007
Recall Number
Z-0532-06
Quantity
3
Firm Location
Tustin, CA

Reason for Recall

In certain conditions the FOV of the lateral plane does not coincide with the X-ray exposed area, and consequently a part of the X-ray field may not be displayed in the magnification mode.

Distribution

CO, MI, & TX

Lot / Code Info

Serial #s W1C0552175, W1C0552181, W1C0562189

Root Cause

Other

Action Taken

An Advisory Letter will be sent to customers via certified mail. Corrected software will be provided at no charge. Customers can contact their local Toshiba Service person to schedule the software update.