Advisor Vital Signs Monitor (model 9200) with Capnography options installed (model 9212S or model 9212A), catalog no. 925654325.

Company
Smiths Medical Pm, Inc.
Recall Initiated
March 16, 2005
Posted
May 12, 2005
Terminated
December 31, 2006
Recall Number
Z-0830-05
Quantity
6
Firm Location
Waukesha, WI

Reason for Recall

Altitude testing of the Advisor with Capnography module found the CO2 (End Tidal and Inspired) values are incorrect when the Advisor is configured to display in the relative units of percent (%) at elevations above 4,100 feet (at standard temperature pressure). The % CO values displayed will be lower than the actual CO2 value.

Distribution

CA and NY within US. Mexico, Argentina, Columbia, Chile, Brazil, Ecuador, Guatemala and Pakistan.

Lot / Code Info

serial no. 706330001, 706330002, 706330003, 706330004, 706330005 and 706431007.

Root Cause

Other

Action Taken

On March 16, 2005, customers were emailed Safety Action Bulletin 05-003 informing them of the low CO2 (End Tidal and Inspired) measurements in percent (%) and urging them to use mmHg or kPa as the unit of measurement . In addition, a Technical Worksheet will be distributed to customers who request a software upgrade. The Technical Worksheet will instruct them how to perform the software upgrade which corrects the CO2 measurements (End Tidal and Inspired) in percent (%).