Roche Cortisol reagent for use in conjunction with Roche cobas e 411, cobas e 601, Elecsys 1010, Elecsys 2010 and Modular E 170 analyzers; Catalog no. 11875116122. in vitro diagnostic.

This recall has been terminated (originally issued June 28, 2007).

Company
Roche Diagnostics Corp.
Recall Initiated
May 15, 2007
Posted
June 28, 2007
Terminated
June 11, 2008
Recall Number
Z-0977-2007
Quantity
6051
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

The test is not as sensitive, or as precise between runs, as claimed in the labeling. Low concentrations are observed only in saliva samples.

Distribution

Nationwide.

Lot / Code Info

Lots 17592001, exp. 10/31/07; 17502002, exp. 10/31/07; 17733201, exp. 2/29/08; and 17733202, exp. 2/29/08.

Root Cause

Other

Action Taken

The firm notified consignees via letter dated 5/15/07 and notifed them of the changes to functional sensitivity and between-run precision claims.