RecallDepth

PROFlex Transporter, Model number 35P. The catalog number for the model 35P is PT3550.

Company
Ferno-Washington Inc
Recall Initiated
March 28, 2003
Posted
November 26, 2005
Terminated
October 29, 2008
Recall Number
Z-0202-06
Quantity
2188 total stretchers
Firm Location
Wilmington, OH

Reason for Recall

The firm received complaints of the stretchers folding. A fold is when the stretcher lowers on its own to either the next position, or any of the other seven postions of the stretcher.

Distribution

The product was distributed nationwide to 27 states: TN, MA, VT, IL, KY, NY, SC, WI, TX, MO, NH, CO, GA, NC, CA, VA, IA, PA, CT, OH, NJ, ID, WV, AL, AR, ME and IN.

Lot / Code Info

All stretchers containing a serial number beginning with the letter ''L'' and followed by a sequence of six numbers.

Root Cause

Other

Action Taken

The firm went out to customers who had complained of a 'fold' and installed the kit on the stretcher that had experienced the fold and on all other PROflexx stretchers owned by the customer. These customers had received a recall letter, dated March 28, 2003. For customers that had not complained of the stretchers 'folding', the firm sent a letter dated June 24, 2003, to all distributors requesting the distributor provide the firm with a list of customers that had received the PROflex stretchers. Once the firm had a list of all customers, they sent them a letter offering them the kit as an enhancement that they could purchase to help reduce the possibility of false locking.