RecallDepth

ACCU-CHEK Inform meter; catalog No. 3035123.

This recall has been terminated (originally issued December 28, 2004).

Company
Roche Diagnostics Corp.
Recall Initiated
November 19, 2004
Posted
December 28, 2004
Terminated
June 7, 2005
Recall Number
Z-0372-05
Quantity
4697
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

A firmware bug may allow test results from other systems, recorded into the INFORM system via the OTE functionality, to be reversed from positive to negative, and vice versa.

Distribution

Nationwide.

Lot / Code Info

All units with firmware version 2.3 that have the OTE (other test entry) functionality enabled.

Root Cause

Other

Action Taken

Consignees were notified via a letter dated 11/19/04, which instructed them to discontinue using the OTE functionality.