Ellik Latex Bulb, Ellik Evacuator. Model Number: 194-2. The Ellik Evacuator and adapter accessories are compatible with Gyrus ACMI Elite and Elite 2 Resection Sheaths.

This recall is currently active, issued July 21, 2025. It was issued by Olympus Corporation Of The Americas.

Company
Olympus Corporation Of The Americas
Recall Initiated
June 18, 2025
Posted
July 21, 2025
Recall Number
Z-2144-2025
Quantity
48 units
Firm Location
Center Valley, PA
Official Source
View on FDA website ↗

Reason for Recall

Discrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).

Distribution

International distribution to the country of Japan.

Lot / Code Info

Model Number: 194-2. UDI: 00821925002593. All lot numbers

Root Cause

Packaging change control

Action Taken

Olympus notified consignees on about 06/18/2025 via letter. Consignees were instructed to review the notification and arrange for the return of all affected units on hand, ensure all affected personnel are aware, and notify customers if the affected units were further distributed. Olympus also requested that consignees complete and return the response form acknowledging receipt of the notification.