Spectrum IQ Infusion Pump, Product Code 3570009

Company
Baxter Healthcare Corporation
Recall Initiated
July 14, 2025
Posted
August 22, 2025
Recall Number
Z-2415-2025
Quantity
6 units
Firm Location
Deerfield, IL

Reason for Recall

Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy.

Distribution

US distribution to states of: AZ, FL

Lot / Code Info

UDI-DI: 00085412610900; Serial Numbers: 3815255 3815261 3815265 3815266 3815267 3815271

Root Cause

Nonconforming Material/Component

Action Taken

Customers were contacted via telephone beginning 7/14/25. The included information about the product, the issue and the action to be taken by the customers. Baxter Healthcare is requesting the return of the impacted product to perform all tests required and confirm that the product perform as expected. If you have any questions, phone (Toll Free): 800-422-9837 or 847-948-4770; hours of operation: Monday through Friday from 8:00 am to 5:00 pm CST