Welch Allyn Life2000 Ventilation system, Product Codes: 1) BT-20-0002; 2) BT-20-0002A; 3) BT-20-0002AP; 4) BT200007; 5) BT-20-0007; 6) BT200011; 7) BT-20-0011; 8) RMS010118CP.

Company
Baxter Healthcare Corporation
Recall Initiated
April 7, 2025
Posted
May 9, 2025
Recall Number
Z-1725-2025
Quantity
4881 units
Firm Location
Deerfield, IL

Reason for Recall

A cybersecurity vulnerability was discovered through internal testing.

Distribution

Worldwide distribution.

Lot / Code Info

ALL SERIAL NUMBERS: 1) BT-20-0002, UDI/DI 00815410020537 or 00887761978089; 2) BT-20-0002A, UDI/DI 00815410020537 of 00887761978089; 3) BT-20-0002AP, UDI/DI 00815410020537 or 00887761978089; 4) BT200007, UDI/DI 00815410020278 or 00887761978089; 5) BT-20-0007, UDI/DI 00815410020278 or 00887761978089; 6) BT200011, UDI/DI 00815410020537 or 00887761978089; 7) BT-20-0011, UDI/DI 00815410020537 or 00887761978089; 8) RMS010118CP, UDI/DI 00815410020537 or 00887761978089.

Root Cause

Software in the Use Environment

Action Taken

Baxter issued an URGENT MEDICAL DEVICE RECALL notice to healthcare providers and patients beginning on 04/07/2025 via USPS first class mail. The notice explained that Baxter was voluntarily initiating a permanent removal of all Life 2000 ventilators currently in use due to a cybersecurity issue and requested the following: Actions to be Taken by DME Providers: 1. Notify your patients using Life2000 ventilators of this recall. 2. Work with patients and prescribers to identify an alternate therapy option. 3. Contact Baxter Acute Care Customer Service team at 800-426-4224, option 2, then option 1, to coordinate return of Life2000 ventilators and compressors. Actions to be Taken by Healthcare Providers: 1. Locate and discontinue use of all Life2000 ventilators and compressors within your facility. 2. Contact Baxter Acute Care Customer Service team at 800-426-4224, option 2, then option 1, to coordinate return of Life2000 ventilators and compressors at your facility. Actions to be Taken by Distributors: 1. Please disseminate this information and conduct a user-level recall of the affected product that you distributed to customers. 2. Contact Baxter Acute Care Customer Service team at 800-426-4224, option 2, then option 1, to coordinate return of Life2000 ventilators and compressors. If you suspect that a device has been exposed to unauthorized personnel, contact Baxter Acute Care Customer Service team. The team can help verify whether the device settings are correct and if the device is performing as expected. The team can be reached by email at HRC_ACCS_Web@baxter.com, or by phone at 800-426-4224, option 2, then option 1. For questions regarding this communication, contact Baxter Acute Care Customer Service team at 800-426-4224, option 2, then option 1, between the hours of 8:00 am and 6:00 pm Central Time, Monday through Friday.

More recalls by Baxter Healthcare Corporation

View all recalls by this company →