MEDLINE ARTERIAL LINE PACK, REF DYNJ65208B, medical convenience kit containing Infusion Catheter with Antimicrobial Heparin Coating.

Company
Medline Industries, Lp - Northfield
Recall Initiated
December 23, 2024
Posted
January 29, 2025
Recall Number
Z-1049-2025
Quantity
50 units
Firm Location
Northfield, IL

Reason for Recall

Certain lots Sterile Infusion Catheters which were mislabeled as having a heparin coating were either distributed or manufactured into medical convenience kits and distributed.

Distribution

US: FL, KS, ND, PA, WA, WV, IL, WA

Lot / Code Info

UDI/DI 10195327561543 (ea) 40195327561544 (cs), Lot Numbers: 24IMA276, 24JMI065, 24FMF777, 24GMG420

Root Cause

Process change control

Action Taken

Medline Industries, LP issued one IMMEDIATE ACTION REQUIRED notice and one MEDICAL DEVICE RECALL notice to its affected consignees on 12/23/2024 via email and US Mail. The IMMEDIATE ACTION REQUIRED notice explained the labeling issue with the device, potential risk, and requested the following actions be taken: 1. affected catheters be quarantined, 2. respond with quantity on hand, 3. if the product was transferred to notify those to whom the affected catheters were transferred. Upon receipt of the response. Medline will issue corrected labeling to affix to the devices. The MEDICAL DEVICE RECALL notice, which pertains to kits manufactured with the affected catheters as a component, also explained the labeling issue with the device, potential risk, and requested the following actions be taken: 1. affected catheters be quarantined, 2. respond with quantity on hand, 3. if the product was transferred to notify those to whom the affected catheters were transferred. Upon receipt of the response. Medline will issue a label to affix to the devices which will state the affected component shall be removed and discarded from further use.

More recalls by Medline Industries, Lp - Northfield

View all recalls by this company →