Discovery MR750w 3.0T, whole body magnetic resonance scanner

This recall is currently active, issued November 22, 2024. It was issued by Ge Medical Systems, Llc.

Company
Ge Medical Systems, Llc
Recall Initiated
October 28, 2024
Posted
November 22, 2024
Recall Number
Z-0517-2025
Quantity
8 units
Firm Location
Waukesha, WI
Official Source
View on FDA website ↗

Reason for Recall

Gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning.

Distribution

Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of: Canada, China, India, Japan, Korea, Latvia, Mexico, Poland, Russia, Saudi Arabia, Spain, and Taiwan.

Lot / Code Info

System ID 760724DVMR1, UDI/DI To be provided; System ID 214820TS750WMR, UDI/DI To be provided; System ID 281333MR2, UDI/DI To be provided; System ID 82427240081, UDI/DI (01)00840682103817(11)181200(21)PG75W1800047SC; System ID 82427120136, UDI/DI Not Applicable; System ID 82427200092, UDI/DI To be provided; System ID 82427250061, UDI/DI To be provided; System ID EM0232, UDI/DI Not Applicable

Root Cause

Device Design

Action Taken

GE Healthcare issued an "URGENT MEDICAL DEVICE CORRECTION" notice to its consignees on 10/28/20224 delivered through a traceable means. The notice explained the issue, safety issue, and requested the following actions be taken for the continued safe usage of the affected devices: "Ensure hearing protection with a Noise Reduction Rating (NRR) of at least 33dB is used during exams on these systems. 200 pairs are provided with this letter. Please post this letter in your facility on or near the MR operator console. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records." Please complete and return the attached acknowledgement form to recall.60998@gehealthcare.com. For questions or concerns, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.