RecallDepth

smiths medical CADD-Solis Li-ion Rechargeable Battery Packs AC adapter, Model Number REF 21-0270-25, for use with the CADD-Solis infusion pumps

Company
Smiths Medical Asd Inc.
Recall Initiated
June 26, 2024
Posted
November 15, 2024
Recall Number
Z-0466-2025
Quantity
156,400 AC adapters
Firm Location
Minneapolis, MN

Reason for Recall

The adapter input plug can become damaged or broken. If the input plug is damaged, the metal contacts to the body of the AC adapter may be exposed or one or more of the AC mains prongs may separate from the input plug.

Distribution

Worldwide distribution.

Lot / Code Info

Model 21-0270-25, UDI/DI 15019517126844, All lots

Root Cause

Device Design

Action Taken

Smiths Medical issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 06/26/2024 via letter distributed by a traceable means. The notice explained the issue, potential risk, and requested the following: Locate all Model 21-0270-25 AC Adapters in your possession. Inspect the AC Adapter for damage to the input plug, missing AC prongs, or exposed metal contacts. If damaged, remove the AC Adapter from service and submit a request for credit to Smiths Medical at 800-258-5361. Distributors were directed to forward to the notice to those to whom the they distributed the device with the request to complete the response form and return it to smithsmedical8528@sedgwick.com.

More recalls by Smiths Medical Asd Inc.

View all recalls by this company →