RecallDepth

1.8mm Truss Wire, REF: 900217, a Component of Metalogix Revolution External Plating System

Company
New Standard Device Inc
Recall Initiated
June 8, 2020
Recall Number
Z-1236-2022
Quantity
426
Firm Location
San Antonio, TX

Reason for Recall

1.8mm truss wires are being removed to facilitate the introduction of 2.0mm truss wires. The increased wire diameter and new manufacturing process are meant to provide a stronger wire and reduced breakage.

Distribution

US Nationwide distribution in the states of AZ, VA, OH, CT, TX, NC, GA.

Lot / Code Info

UDI: 854641008419, Lots: A29518X and A17219DA

Root Cause

Device Design

Action Taken

On 06/08/2020, Market Withdrawal notices were emailed to customers who were asked to remove affected truss wires from inventory and return to the recalling firm. Customers were informed to use alternate 1.8mm and 2.0mm unaffected smooth wires. Upon completion of verification testing, both 2.0mm truss and smooth wires will be released for use. Customers with additional questions are encouraged to call (833) 659-2019 Ext 6.

More recalls by New Standard Device Inc