RecallDepth

Metalogix Self-Drill Half-Pin, a component of the Revolution External Fixation System. Orthopedic. Metalogix Self-Drill Half-Pin 5x30x215mm REF 620030 Metalogix Self-Drill Half-Pin 5x40x215mm REF 620040 Metalogix Self-Drill Half-Pin 5x60x215mm REF 620060 Metalogix Self-Drill Half-Pin 5x80x215mm REF 620080

This recall is currently active, issued December 28, 2020. It was issued by New Standard Device Inc.

Company
New Standard Device Inc
Recall Initiated
December 28, 2020
Recall Number
Z-1649-2022
Quantity
410 total devices
Firm Location
San Antonio, TX
Official Source
View on FDA website ↗

Reason for Recall

Complaints were received for Self-drill Half-Pin tip breakages occurring during surgery.

Distribution

U.S.: AZ, GA, IL MT, NC, NJ, NV, OH, PA, TX, and VA O.U.S.: None

Lot / Code Info

Part Number / DESCRIPTION / UDI-DI Code / Lot Numbers: 620030 / Self-Drill Half-Pin 5x30x215mm / 854641008198 / A29418X, A17219CA 620040 / Self-Drill Half-Pin 5x40x215mm / 854641008204 / A29418Y, A17219CB 620060 / Self-Drill Half-Pin 5x60x215mm / 854641008211 / A29418Z, A17219CC 620080 / Self-Drill Half-Pin 5x80x215mm / 854641008228 / A29518A, A17219CD

Root Cause

Device Design

Action Taken

On 12/28/2020, Metalogix send a "Field Notice: Self-Drill Half-Pin" via email. The Notice only informed customer that the Citieffe Half-Pins will temporarily be used to replace the current 5mm Self-Drill Half-Pins. A contact number for Mr. Elton Ely, QA Manager was provided in the notice 833-659-2019 ext. 6

More recalls by New Standard Device Inc

View all recalls by this company →