Medline procedure packs containing Shenli syringes, labeled as follows: 1) CVC INSERT BUNDLE 16CM, Pack Number ECVC3260; 2) VANTEX 7FR 3L 20CM CVC BUNDLE US, Pack Number ECVC8045
This recall is currently active, issued August 27, 2024. It was issued by Medline Industries, Lp - Northfield.
- Recall Initiated
- April 8, 2024
- Posted
- August 27, 2024
- Recall Number
- Z-2629-2024
- Quantity
- 1,679,067 units in total
- Firm Location
- Northfield, IL
- Official Source
- View on FDA website ↗
Reason for Recall
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Distribution
Worldwide distribution.
Lot / Code Info
1) ECVC3260, Lot Number 2018052950; 2) ECVC8045, Lot Number 2023110290
Root Cause
Nonconforming Material/Component
Action Taken
Medline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request these syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods.