Phantom Nail System, Drill, 4.6 x 300mm, Cannulated, 3/16" Sq. Connection p28 P99-110-4630

Company
Paragon 28, Inc. 14445 Grasslands Dr Englewood Co 80112-7062
Recall Initiated
July 30, 2021
Terminated
December 15, 2021
Recall Number
Z-2488-2021
Quantity
14 drill bits

Reason for Recall

A thin wall condition between the inner cannulation and the flutes of the drill which potentially could cause a fracture through the wall of the drill to the cannulation beneath.

Distribution

U.S. Nationwide distribution in the states of CA, FL, GA, ID, KS, KY, LA, MI, MO, NC, NY, TX, VA, WA and WV. O.U.S.: N/A

Lot / Code Info

Model Number: P99-110-4630 UDI Code: (01)00889795056849(10)HN120550 Lot Number: HN120550

Root Cause

Process change control

Action Taken

On 08/12/2021, the Recalling Firm emailed to its distributors an "Medical Device Recall" Notification Letter inform them that it was identified that there is a thin wall condition between the inner cannulation and the flutes of the drill. At worst case, the defect is a fracture through the wall of the drill to the cannulation beneath. This feature is potentially nonconforming in some units within the lot identified below. The distributor was instructed to: 1) Immediately check its inventory and hospitals/health care facility within their region and to identify, discontinue use, quarantine and not distribute any further the affected products. 2) Complete the Medical Device Recall Response Form and return to mkinnee@paragon28.com. The form is to be completed and returned even if no physical product is in their possession. 3) Immediately return affected devices along with a copy of the Medical Device Recall Response Form to the Recalling Firm at Paragon28, Attn: Recalls, RN-008, 14445 Grasslands Dr., Englewood, CO 80112. For questions or further assistance - contact the Recalling Firm at: email - mkinnee@paragon28.com or complaints@paragon28.com Phone - 949-280-7688

More recalls by Paragon 28, Inc. 14445 Grasslands Dr Englewood Co 80112-7062

View all recalls by this company →