Phantom Fibula Nail, 2.8mm x 130mm, Right, Ti, Fibula Nail, REF: P36-128-130R-S

Company
Paragon 28, Inc. 14445 Grasslands Dr Englewood Co 80112-7062
Recall Initiated
October 13, 2025
Posted
December 2, 2025
Recall Number
Z-0871-2026
Quantity
20

Reason for Recall

Fibula nail system intended for use in fixation of fibular fractures and osteotomies, has out of specification thread depths on the fibula nail, so implant may not be able to mate properly with the associated mounting bolt. If nail (implant) cannot properly mate with instrumentation, the surgery potentially will need to be aborted.

Distribution

US Nationwide distribution in the states of NJ, WA, NY, MD, CA, FL, NM, TX, IL, OH, NV, TN, LA.

Lot / Code Info

UDI-DI: 00889795133670. Lot: 501182207125. Expiration: March 13th, 2030

Root Cause

Device Design

Action Taken

On 10/13/2025, recall notices were emailed to distributors who were asked to do the following: 1) Return affected devices to the firm using return label. 2) Ensure that in instances where you have distributed this affected product, that you forward this Recall Notification to your customers. 3) Replacements are available by contacting Orders@paragon28.com Complaints can be directed to complaints@paragon28.com