LigaSure" Exact Dissector Nano Coated 20.6mm-21cm Product Usage: A bipolar electrosurgical instrument intended for use in open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired.

This recall has been terminated (originally issued March 15, 2018).

Company
Covidien Llc
Recall Initiated
March 15, 2018
Terminated
December 11, 2019
Recall Number
Z-2427-2018
Quantity
936
Firm Location
North Haven, CT
Official Source
View on FDA website ↗

Reason for Recall

Devices were found with a sterile breach (puncture) in the pouch from a sort of 8,064 packaged devices.

Distribution

Worldwide Distribution in the countries of Australia, Austria, Belgium, Denmark, Finland, France, Germany, Ireland, Italy, Netherlands, Norway, Poland, Saudi Arabia, Spain, Swede, Switzerland and United Kingdom

Lot / Code Info

Lot #s 72550127X 72770140X 72770141X 73000048X

Root Cause

Other

Action Taken

Medtronic Regional sent a notification letter to customers via overnight delivery. the letter identified the affected product, problem and actions to be taken. The letter request that you quarantine and return any unused products of the affected item codes and production lots. Unused products from the affected item codes should be returned. If you have distributed Covidien LigaSure" Exact Dissector listed, please promptly forward the information from this letter to those recipients. All unused products from the affected item codes must be returned. The potentially affected product was shipped to your facility between November 2017 and February 2018.