RecallDepth

TRACH PACK. Sterile Procedural Tray.

Company
Medline Industries, Lp - Northfield
Recall Initiated
November 17, 2022
Posted
February 6, 2023
Recall Number
Z-1061-2023
Quantity
28 cases (112 units)
Firm Location
Northfield, IL

Reason for Recall

Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Japan, Mexico, Panama, Turks & Caicos, United Arab Emirates.

Lot / Code Info

Item Number: DYNJ46518G, Case UDI/GTIN: 40195327009169, Unit UDI/GTIN: 10195327009168, Lot Number: 22JMB321, 22JMA904, 22HMD287

Root Cause

Device Design

Action Taken

Medline Industries, LP notified consignees via email and first-class mail on 11/15/2022. Consignees were instructed to check stock for affected product and immediately quarantine. For any affected units, customers were instructed to over-label the sterile kits with a warning label indicating "Warning To Avoid Injury, Please Handle Blades with Care." No product is to be returned. Customers were also instructed to complete and return the response form and notify customers if further distributed.

More recalls by Medline Industries, Lp - Northfield

View all recalls by this company →