Philips DigiTrak XT Holter Recorder Software version number B.1 and C.2. The DigiTrak XT Recorder is available in four different configurations. Service tag numbers for these configurations are as follows: - 453564067151 24-Hour Configuration - 453564067161 48-Hour Configuration - 453564067171 96-Hour Configuration - 453564067181 7-Day Configuration Model Number: 860322

This recall has been terminated (originally issued January 13, 2020).

Company
Philips North America, Llc
Recall Initiated
January 13, 2020
Terminated
June 22, 2023
Recall Number
Z-1345-2020
Quantity
44,442 units
Firm Location
Andover, MA
Official Source
View on FDA website ↗

Reason for Recall

if an AAA battery is inserted in the recorder and a user attempts to start it, or if the recorder is inserted in the docking station, the recorder will display Error: 602 and fail to function for a new patient study.

Distribution

US Nationwide Foreign: Afghanistan Albania Algeria Argentina Armenia Aruba Australia Austria Azerbaijan Bangladesh Belarus Belgium Bermuda Bulgaria Burkina Faso Canada China C¿te D'Ivoire Croatia Curacao Cyprus Czech Republic Egypt France Germany Ghana Gibraltar Greece Guatemala Hong Kong Hungary Iceland India Indonesia Iran Iraq Ireland Israel Italy Japan Jordan Kenya Korea, Republic of Kosovo Kuwait Latvia Lebanon Libya Luxembourg Macao Macedonia Malaysia Maldives Mali Malta Mauritius Mexico Mongolia Myanmar Nepal Netherlands Netherlands Antilles New Zealand Nicaragua Nigeria Norway Oman Pakistan Palestine, State of Panama Peru Philippines Poland Portugal Puerto Rico Qatar Romania Russian Federation Rwanda Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Suriname Sweden Switzerland Taiwan Thailand Trinidad and Tobago Turkey Uganda United Arab Emirates United Kingdom Uzbekistan Venezuela Viet Nam

Lot / Code Info

All Philips 860322 DigiTrak XT Holter Recorder, manufactured or distributed before December 31, 2019, are affected, regardless of serial number, ship date, and date of manufacture.

Root Cause

Device Design

Action Taken

Philips Issued Medical Device Correction letter dated January 13, 2020 to U.S. customers, via priority mail. Philips Market Organizations are responsible for distributing the letters outside of the U.S. Instructions to Customer Philips is asking customers to follow the Action to be Taken by Customer/User section of the CIL: Upon receipt of this notification, promptly locate and isolate affected devices. The steps in the Instructions on Clearing Error 602 below may be used to reset your devices. Customers can contact Philips should they choose not to reset the device on their own. For information or support contact: your local Philips representative or call us at 1 -800-722-9377.