RecallDepth

Perfusion Pack, Rx only, Sterile EO

Company
Livanova Usa
Recall Initiated
November 9, 2018
Terminated
May 3, 2022
Recall Number
Z-0770-2019
Quantity
483
Firm Location
Arvada, CO

Reason for Recall

4-way stopcocks and stopcock manifold assembled in certain perfusion tubing systems, indicated for use in surgical procedures requiring extracorporeal support for periods of up to six hours, may leak or be occluded.

Distribution

Worldwide Distribution - Domestic (US) Consignees: AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WV, MS; Foreign (OUS) Consignees: Canada

Lot / Code Info

Lot Numbers: 1819000027, 1813500044, 1809300053, 1817900004, 1821900089, 1812100018, 1816200009, 1817600016, 1810600004, 1811400014, 1812000017, 1812700017, 1813500013, 1815500011, 1816400023, 1809400005, 1811600025, 1812000018, 1812700018, 1813500014, 1814100012, 1815800010, 1810600006, 1811300003, 1812000019, 1812700020, 1815500012, 1817600017, 1813400007, 1814200096, 1816200011, 1817100013, 1813600009, 1817100103, 1819100014, 1810600014, 1812000030, 1815600027, 1817200017, 1819100017, 1810600015, 1816300051, 1815800015, 1817800028, 1810100021, 1810800022, 1811500019, 1812200022, 1813500020, 1814400026, 1815800017, 1816500016, 1817900015, 1810000025, 1810800023, 1811400022, 1812200023, 1813400010, 1814100019, 1816300053, 1810100022, 1812000034, 1812700032, 1814400027, 1816400032, 1817800031, 1819200020, 1812700034, 1814900020, 1817600031, 1814300026, 1818400082, 1819700023, 1820400020, 1814100022, 1816400035, 1818600024, 1811000017, 1815100030, 1816400036, 1819000012, 1809500011, 1810700016, 1812000036, 1814100023, 1815800018, 1810200011, 1814900021, 1817900017, 1806000020, 1807200053, 1810700018, 1811300015, 1812700035, 1816400037, 1819000013, 1810000027, 1812800072, 1814900022, 1813700002, 1817700022, 1819200022, 1810100026, 1811400028, 1803900021, 1805300025, 1805900033, 1806600023, 1807100016, 1810700020, 1811500023, 1813700014, 1814400028, 1816200018, 1817200020, 1811600032, 1814900023, 1816500019, 1817900019, 1812800073, 1815800019, 1819000016, 1810600022, 1813600016, 1818300026, 1810200014, 1812200029, 1814200103, 1815600031, 1816400039, 1818400085, 1821100028, 1810600023, 1810900022, 1813000021, 1817900021, 1803600019, 1803800017, 1805900035, 1806100020, 1806700039, 1807400038, 1808600021, 1812900030, 1813000022, 1806600064, 1814400031, 1816500022, 1818300027, 1817200022, 1810800024, 1812300012, 1804500029, 1806000024, 1810600024, 1812300013, 1818300028, 1810800025, 1812300014, 1813600019, 1819100023, 1811400001, 1811300019, 1811600034, 1812200032, 1814400033, 1819000020, 1809400017, 1814900027, 1815600032, 1810600026, 1811600035, 1817700026, 1819200027, 1810100028, 1813400015, 1815800021, 1817600036, 1810200016, 1811400031, 1812700040, 1815500028, 1816200021, 1816500024, 1818600028, 1811500024, 1812900031, 1814400034, 1816900018, 1812900032, 1817100021, 1817800035, 1813000025, 1816300058, 1819200028, 1811400033, 1814200105, 1816900020, 1817900023, 1807800020, 1808100036, 1805200067, 1810100030, 1819200029, 1806000029, 1807400042, 1813400017, 1816200024, 1815000041, 1818600030, 1819100025, 1810800030, 1812300017, 1816200027, 1816400043, 1817000037, 1817600038, 1819300025, 1809300068, 1813600022, 1816400079, 1821100036, 1809500017, 1811500029, 1811600040, 1812200037, 1814300030, 1815100034, 1815800023, 1817000038, 1817700032, 1818600032, 1812100075, 1813400057, 1814200109, 1814300031, 1820600034, 1810600029, 1812900034, 1813500027, 1819000023, 1810800033, 1813000032, 1813700022, 1816500025, 1814400069, 1810600032, 1811500033, 1813400021, 1815100037, 1818400090, 1812900039, 1813500029, 1816500026, 1818600034, 1810800034, 1811300023, 1812000043, 1813700025, 1814900031, 1816200029, 1816500027, 1817000040, 1818300031, 1804600025, 1805800065, 1806600028, 1810100035, 1811600042, 1813500030, 1816200030, 1812200001, 1816400087, 1816400048, 1817600041, 1811000008, 1810800065, 1817100025, 1817600042, 1817900029, 1812000088, 1818600079, 1816400050, 1812900041, 1814100030, 1815800024, 1810100081, 1810700112, 1811300062, 1817600043, 1818300033, 1805200031, 1805900040, 1807200067, 1812100042, 1815000047, 1817100027, 1811500035, 1812800074, 1822100076, 1810900029, 1813000033, 1814200112, 1814900035, 1815600039, 1816400051, 1814100032, 1816900025, 1812200045, 1815600041, 1817000042, 1817900030, 1812000010, 1805200035, 1806600030, 1809500023, 1811400037, 1813600030, 1810000033, 1811600046, 1813000039, 1815100041, 1815800028, 1817100030, 1818400143, 1814900036, 1813600001, 1814100001, 1812300001, 1815100065, 1809500025, 1811400040, 1812200050, 1813500036, 1813700027, 1815800031, 1816500032, 1818600039, 1810600036, 1811500039, 1812900044, 1816200031, 1817000044, 1818400096, 1817700085, 1818600075, 1809900043, 1811600049, 1817700038, 1818600040, 1807100026, 1807800034, 1810000035, 1812300023, 1810700030, 1809900044, 1812200051, 1819000029, 1819200036, 1810600037, 1811500042, 1812200052, 1814900040, 1816200033, 1817800043, 1815800033, 1821200047, 1809500028, 1812000047, 1809900048, 1810800040, 1813500041, 1815500034, 1817700039, 1816500033, 1805100104, 1806400043, 1807400052, 1809500030, 1812100053, 1813500042, 1818400097, 1810200032, 1815600050, 1816500036, 1810800043, 1813400028, 1818300039, 1818400098, 1815600051, 1810800044, 1818300040, 1810200033, 1812300026, 1815100046, 1817800070, 1818600042, 1810100043, 1817200037, 1812000050, 1812300029, 1816300076, 1817700042, 1810100044, 1810700034, 1811400048, 1812300030, 1813700031, 1817200038, 1807100034, 1803700065, 1806400048, 1807100035, 1808500045, 1810800045, 1812000051, 1814300043, 1815700038, 1817700043, 1818600044, 1803700066, 1805200043, 1811400062, 1813400034, 1813700032, 1810700038, 1803600044, 1806100041, 1807100040, 1808800042, 1812300032, 1813600036, 1816200040, 1817900042, 1810800046, 1811500052, 1812800085, 1813600037, 1818600047, 1803700072, 1811600060, 1812700059, 1815800045, 1819100045, 1813600039, 1815800046, 1817800051, 1808600043, 1803600047, 1803700075, 1811000043, 1812700061, 1814400048, 1817800052, 1809500037, 1811000044, 1815500042, 1816400064, 1818600051, 1810000045, 1812000062, 1811300042, 1811600066, 1815500045, 1819000047, 1816400068, 1817100047, 1819100050, 1809500040, 1811600067, 1813500049, 1815000056, 1817800054, 1812000002, 1813700038, 1815800050, 1806400055, 1808500048, 1806600059, 1819000051, 1817800055, 1813400049, 1819000054

Root Cause

Process control

Action Taken

On 11/09/2018 Medical Device Correction Notices were mailed to customers via UPS First Overnight. Customers were advised: 1) Inspect your stock and identify Perfusion Tubing System(s) belonging to the impacted lots in inventory, as noted in Attachment 1. 2) The recalling firm will be shipping one case of sterile stopcocks (part number: SCO4S) and manifolds (part number: 020581600) to your facility as an immediate replacement for those remaining that may be defective in your impacted Perfusion Tubing System stock. 3) As each affected pack is opened for use, please replace the original stopcocks in the pack with the spare provided by the firm. The original stopcocks should be destroyed/ discarded. 4) Contact your firm Representative if additional replacement stopcocks are required. In addition, customers were asked to complete and return the attached Customer Response Form. Also, to assure within their organization that the correction notice is communicated to all personnel who need to be aware of it. For questions about this Medical Device Correction, please contact (800) 986-4702 or e-mail USFSN@livanova.com

More recalls by Livanova Usa

View all recalls by this company →