Sorin Stockert Heater-Cooler System 3T, a) Item #: 16-02-81, 240 v/60 Hz; b) Item #: 16-02-82, 208 v/60 Hz; c) Item #: 16-02-85, 120 v/60 Hz

Company
Livanova Usa
Recall Initiated
April 20, 2018
Posted
May 18, 2018
Terminated
February 5, 2025
Recall Number
Z-1901-2018
Quantity
2408
Firm Location
Arvada, CO

Reason for Recall

Firm is notifying customers that the Deep-Cleaning Service is now available in the United States. Stockert Heater-Cooler System 3T devices are suspected of microbial contamination and may contain visible biofilm.

Distribution

MN, TX, WA, OR, IL, OH, SC, AK, PA, CA, WI, GA, NJ, FL, AZ, KY, MS, MO, LA, AR, MA, MI, MT, CO, TN, NE, NC, VA, NV, HI, PR, WY, AL, OK, IN, CT, ME, IA, SC, KS, NM, SD, DC, MD, ND,

Lot / Code Info

Serial #: a) 16S10743 - 6S11708  b) 16S10772 - 16S16840  C) 16S10958 - 16S16847

Root Cause

Device Design

Action Taken

On 04/20/2018, a Medical Device Correction notice was sent to customers informing them to return devices less than ten years old. Returned devices will receive a deep cleaning service, which will reduce microbial contamination to HPC < 100 CFU/ml of water and NTM < 1 CFU/100ml.