smith&nephew TRIGEN INTERTAN 10S, REF 71675384, 10 MM X 18 CM NAIL, 130o (degrees), SET SCREW PREASSEMBLED IN NAIL

Company
Smith & Nephew Inc
Recall Initiated
October 24, 2017
Recall Number
Z-0097-2018
Quantity
17 units
Firm Location
Memphis, TN

Reason for Recall

A single lot of INTERTAN 10S 10 MM X 18 CM 130 D was manufactured with the screw inserted upside down.

Distribution

US Distribution - TN, IL, CT, MO, CA., and Internationally to Italy and Germany

Lot / Code Info

Lot number 16LM07111

Root Cause

Employee error

Action Taken

Smith & Nephew sent an Urgent Medical Device Recall Notice dated October 24, 2017, to all affected customers. The firm requested return of the product. Please follow the instructions on the attached Response Form. For further questions, please call (978) 749-1440.