Leica HistoCore SPECTRA CV Coverslipper, used in preparation of histological and cytological tissue samples for microscopic analysis for medical diagnostics.

This recall has been terminated (originally issued September 4, 2018).

Company
Leica Microsystems, Inc.
Recall Initiated
September 4, 2018
Terminated
July 10, 2020
Recall Number
Z-0019-2019
Quantity
1 unit
Firm Location
Buffalo Grove, IL
Official Source
View on FDA website ↗

Reason for Recall

Product was assembled with an isolating fiberglass hose within the oven in the wrong position. This results in improper shielding from potential electrical shock.

Distribution

Distribution in US state of MO.

Lot / Code Info

Serial Number 102

Root Cause

Process design

Action Taken

The firm issued a Field Action Notice by letter on 09/05/2018. The notice explained the issue and the possible hazard. The firm will schedule a visit with a service engineer to perform the correction.