MEVION S250 for Proton Radiation Therapy

Company
Mevion Medical Systems, Inc.
Recall Initiated
March 15, 2016
Posted
April 20, 2016
Terminated
August 24, 2017
Recall Number
Z-1529-2016
Quantity
5
Firm Location
Littleton, MA

Reason for Recall

Possibility of Gantry Motion with Aperture or Compensator only partially inserted. An Aperture may fall out during Gantry motion. It is possible that it can hit a patient and cause traumatic injury or death.

Distribution

US nationwide distribution to MO, OK, FL, and NJ.

Lot / Code Info

S250-0002

Root Cause

Device Design

Action Taken

Customer notification letters were sent to customers on March 15, 2016, by certified mail, return receipt requested, as well as by email. The letter explains the problem, description of potential harm, and the cause containment, and mitigation. The cause of the malfunctioning sensors was discovered to be a non-conforming limit switch assembly. A Mevion Field Service Engineer will immediately inspect all such parts in the field for similar non-conformance, and replace those parts if necessary. In addition, a new periodic preventive maintenance check will be performed to individually test each of the redundant sensors at all sites. In the interim period, on a daily basis, until the inspection, repair, and the preventive maintenance check is completed at any site, Mevion Field Service will check that the Applicator Door interlock sensors are working properly. Mevion will also update the inspection procedure for incoming parts to ensure that any non-conformance is identified early in the manufacturing process.

More recalls by Mevion Medical Systems, Inc.

View all recalls by this company →