MEVION S250 Proton Radiation Treatment System; intended to deliver proton radiation treatment to patients

Company
Mevion Medical Systems, Inc.
Recall Initiated
June 23, 2020
Terminated
December 9, 2020
Recall Number
Z-2686-2020
Quantity
8 units
Firm Location
Littleton, MA

Reason for Recall

The perforated screen that divides the inside of the process water tank has a very sharp edge across the top of the Heat Exchanger leading to potential injury for servicer.

Distribution

MO, NJ, OK, FL, OH, DC, Netherlands

Lot / Code Info

Lot #s: MEVION S250-0001 MEVION S250-0002 MEVION S250-0003 MEVION S250-0004 MEVION S250-0005 MEVION S250-0006 MEVION S250-0007 MEVION S250-0008  UDI: (01)00864366000100 (11) 131209 (01)00864366000100 (11)150526 (01)00864366000100 (11)150113 (01)00864366000100 (11)150430 (01)00864366000100 (11)160127 (01)00864366000100 (11)160603 (01)00864366000124 (11) 180220 (01)00864366000124 (11) 181219

Root Cause

Device Design

Action Taken

This screen would not be accessed by normal user (only servicer). Firm is notifying servicing personnel and not consumer/user level. No customer letters/communication sent. Servicer will install edge guard on effected units. Technical Service Bulletin issued to servicer.