TSX-IOI: AQUILION 32/64; System, X-Ray, Tomography, Computed. Designed to produce cross-sectional images of a human body.

This recall has been terminated (originally issued April 8, 2013).

Company
Toshiba American Medical Systems Inc
Recall Initiated
November 30, 2011
Posted
April 8, 2013
Terminated
June 17, 2013
Recall Number
Z-1067-2013
Quantity
23 units
Firm Location
Tustin, CA
Official Source
View on FDA website ↗

Reason for Recall

The firm initiated this recall due to a potential software issue. The DLP (Dose Length Product) value may be displayed incorrectly when the vHP option is used.

Distribution

Nationwide Distribution including the states of WA, OH, CA, TX, WV, IL, AZ, CA, WI, ME, WI, NC, FL, MT, and HI.

Lot / Code Info

Serial Numbers: HDA0722595 JDA0972130 JDA0972135 JDA0992147 JDA09Y2158 KDA09X2063 JDA09Y2157 JDA09Y2161 JDA1012171 JDA10X2191 JDA10X2197 JDA10Z2214 JGAl112220 JLA1122232 JLA1132235 JDAl112216 JDAl112219 NLA0972032 JDA1182242 JDA1182241 JDA1182244 JDA1072183 JDA1082189

Root Cause

Process design

Action Taken

Toshiba sent "URGENT: MEDICAL DEVICE CORRECTION" letters to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. The letter contained a return reply form that is to be faxed to TAMS for retention. Contact the firm at (800) 421-1968 for questions relating to this letter.