I-STAT EG7+ Cartridge Abbott Point of Care, Princeton, NJ. The i-Stat EG7 cartridge is intended to be used for quantitatively measure sodium, potassium, ionized calcium, oxygen, carbon dioxide, partial pressure in arterial, venous, or capillary whole blood.

This recall has been terminated (originally issued August 2, 2012).

Company
Abbott Point Of Care Inc.
Recall Initiated
January 1, 2011
Posted
August 2, 2012
Terminated
May 22, 2013
Recall Number
Z-2133-2012
Quantity
52,750 cartridges
Firm Location
Princeton, NJ
Official Source
View on FDA website ↗

Reason for Recall

i-Stat EG7 cartridges may generate falsely depressed PCO2 results for samples above 26 mm hg when stored at room temperature for four weeks or longer.

Distribution

USA ( nationwide )

Lot / Code Info

List number 03P76-25 Lot number S10246A

Root Cause

Nonconforming Material/Component

Action Taken

Abbott Point of Care, Inc sent a Urgent Recall Notice letter dated January 14, 2011 to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to discontinue use of Cartridges from the identified lot. All unused cartridges will be credited and are to be returned as instructed on the enclosed Business Reply Card. If you have any questions regarding this information please contact your Abbott Point of Care Technical Support at 800-366-8020 Option 1, or your Abbott Point of Care representative.