ARTISTE , ONCOR and PRIMUS accelerators that utilize the RT Therapist v4.2 system in combination with syngo RT Oncologist 4.2. The ONCOR and PRIMUS may be updated to utilize the RT Therapist v4.2 software. The RT Oncologist v4.2 is currently only on Non-US systems. Distributed by Siemens Healthcare, Concord, CA 94520 Intended use: Radiation therapy

This recall has been terminated (originally issued May 19, 2011).

Company
Siemens Medical Solutions Usa, Inc
Recall Initiated
March 31, 2011
Posted
May 19, 2011
Terminated
March 30, 2012
Recall Number
Z-2216-2011
Quantity
45 active devices
Firm Location
Concord, CA
Official Source
View on FDA website ↗

Reason for Recall

Siemens has become aware of a potential safety issue that may result in mistreatment when customers are configured with the ARTISTE Linac RT Therapist v4.2 in combination with RT Oncologist v.4.2 and use the remote "Online 2D Image Review.

Distribution

Worldwide distribution: USA (nationwide) including states of: CA, FL, OH, PA and UT; and countries including: Australia, Belgium, Croatia, Germany, Japan, South Africa, Sweden and United Kingdom.

Lot / Code Info

Therapist Exoress Basis Part number 08515289 syngo Therapist Express Part Number 08162815 syngo RT Therapist Connect/ MOSAIQ OIS Part number 08168754.

Root Cause

Software design

Action Taken

The firm, Siemens, sent an "Urgent: Medical Device Correction Customer Safety Advisory Notice" on March 31, 2011 via Siemens Customer Service Group or certified mail to its customers. The notice describes the product, problem and actions to be taken. The customers were instructed to not use the "Online 2D Image Review" on the Syngo RT Oncologist 4.2; to always calculate the offset directly on the RTT 4.2 before it is applied for the treatment with the patient on the treatment table; include this notice in their syngo RT Therapist 4.2 System Owner Manual chapter "Safety Advisory Letter" where it should remain until the measures defined in the notice are performed, and in the interest of safety, perform the preventative measures noted in this notice and inform all affected personnel immediately. Siemens is currently working on a solution for the incremental loading workflow in syngo RT Therapist 4.2 and will provide a software update as soon as possible. Siemens will notify the customers as soon as an update for their system will be available. UPDATE: The release of Safety Update UI TH010/11/S to distribute the software update for RTTv4.2.94O04 began on April 15, 2011. If you have any questions, contact Regulatory Affairs at 925-602-8157.