Metha(R) Hip System Cobalt Chrome (CoCr) Metha Modular Neck Adapters used with the Metha Modular Short Hip Stems. Aesculap AG. Intended to replace a hip joint

Company
Aesculap Inc
Recall Initiated
July 19, 2011
Posted
August 12, 2011
Terminated
March 19, 2013
Recall Number
Z-3014-2011
Quantity
149
Firm Location
Center Valley, PA

Reason for Recall

Aesculap Implant Systems received a notification from the manufacturer, Aesculap, AG of a fractured CoCr neck adapter.

Distribution

Nationwide (USA) Distribution including the states of AL, FL, GA, MD, NM, NY, OK, and TN.

Lot / Code Info

Part Numbers: NC077K, NC078K, NC079K, NC087K, NC088K, NC089K, NC097K, NC098K, and NC099K)  Lot Numbers: All distributed lot numbers.

Root Cause

Unknown/Undetermined by firm

Action Taken

AESCULAP Implant Systems sent an "Important Recall Notification" letter dated to all affected customers. The informed customers of a complaint received where fractured CoCR adapter was found and to return any affected product to the firm. The reason for the fracture is currently under investigation. While the firm is trying to determine the root cause of the fracture, they will be offering the Metha non-modular short hip implants to their customers. Customers are asked to return an Inventory Sheet to AESCULAP. For additional information on this recall call (610) 984-9239 or (610) 984-9300.