Baxter Solution Set with Duo-Vent Spike; 109" (2.8 m), Male Luer Lock Adapter 10 drops/mL. Baxter Healthcare Corporation. A sterile, single use, nonpyrogenic fluid pathway for use with rigid nonvented and flexible containers.

Company
Baxter Healthcare Corp. Rt
Recall Initiated
July 21, 2011
Posted
August 3, 2011
Terminated
June 18, 2012
Recall Number
Z-2920-2011
Quantity
4,752 units
Firm Location
Round Lake, IL

Reason for Recall

The Solution Sets are being recalled because the position of the slide clamp and roller clamp may have been reversed during assembly. The reversed orientation of the slide clamp and roller clamp will cause the set to be incorrectly loaded into an infusion pump, resulting in fluid flow from the patient to the pump, rather than from the pump to the patient.

Distribution

(USA) Nationwide Distribution including the states of Alabama, California, Iowa, Illinois, Indiana, Massachusetts, Maryland, Michigan, North Carolina, New Hampshire, Ohio, Pennsylvania, Tennessee and Virginia.

Lot / Code Info

Product Code 1C8507s, lot numbers GR291542, GR292409, GR292797 and GR294454.

Root Cause

Employee error

Action Taken

Baxter Healthcare Corporation sent a "Urgent Product Recall" letter dated July 21, 2011 to all affected customers. The letter included a description of the problem and requested they contact Baxter at 1-888-229-0001 to return all affected product. They were asked to complete and return the attached customer reply form, and to contact and notify their customers. For questions contact Baxter at 1-800-422-9837 during the hours of 8:00 am to 5:00 pm CST.