RecallDepth

KIMBERLY CLARK MIC GASTRO-ENTERIC FEEDING TUBE, 22F. Gastro-Enteric Feeding Tube.

This recall has been terminated (originally issued February 2, 2012).

Company
Kimberly-Clark Corporation
Recall Initiated
December 21, 2011
Posted
February 2, 2012
Terminated
July 2, 2012
Recall Number
Z-0938-2012
Quantity
50 devices
Firm Location
Roswell, GA
Official Source
View on FDA website ↗

Reason for Recall

The product is misbranded, i.e. the Gastric (center) port may be mislabeled as the Jejunal port, and the Jejunal (side) port mislabeled as the Gastric port.

Distribution

Worldwide Distribution-USA (nationwide) including the states of CA, CO, ID, IL, MD, MO, NC, NJ, NY, PA, SC, UT, VA, WA, WI, and WV and the country of Canada.

Lot / Code Info

Catalogue number: 0210-22, Lot number: AA1129E02

Root Cause

Labeling False and Misleading

Action Taken

Kimberly-Clark Health Care sent an "URGENT: VOLUNTARY MEDICAL DEVICE RECALL NOTIFICATION" letter and e-mail to all affected customers. The communications described the product, problem, and actions to be taken by the customers. Customers were instructed to locate and quarantine any affected product. A Product Recall Response Sheet was enclosed for customers to complete and return to the firm via fax at 920-380-6682. Contact the firm at 770-587-7400 for questions regarding this recall.

More recalls by Kimberly-Clark Corporation

View all recalls by this company →