ECMO Heater Model 333, Cardiopulmonary bypass temperature controller. This device is intended to provide temperature-controlled warm water flow to a blood heat exchanger in order to maintain a patient's blood temperature. It can be used during extracorporeal circulation procedures. It is intended to use with commercially available heat exchangers.
- Company
- Cincinnati Sub-Zero Products Inc
- Recall Initiated
- May 2, 2011
- Posted
- July 1, 2011
- Terminated
- March 7, 2013
- Recall Number
- Z-2720-2011
- Quantity
- 116 units
- Firm Location
- Cincinnati, OH
Reason for Recall
The probe jack contacts (terminals) on certain of the firm's ECMO Heater devices are located too close to the solid state relay contacts (terminals) which can lead to the terminals coming into contact with each other. If this situation occurs, the service technician, user, and/or patient could be exposed to electric shock during servicing and/or use.
Distribution
Worldwide Distribution and countries of The Netherlands, Kuwait, India, Saudi Arabia, Taiwan and Australia. (USA) Nationwide including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IL, IN, LA, MA, MD, MI, MIN, MO, NC, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, WA, and WI.
Lot / Code Info
Model: 333 Part Number: 86130 Serial number: 864-E1180 through 924-E2528
Root Cause
Device Design
Action Taken
Cincinnati Sub-Zero Products Inc issued an "Urgeant Medical Device Field Action" letter on May 2, 2011 to all affected customers. The letter includes; description of product and problem, affected lots and instructs customers to discontinue its use. The product is to be returned to the Service Department for an update and the customer is instructed to complete and mail Field Action Form. For additional contact the firm at 1-800-989-7373 or (513) 772-8810.