Cold Therapy Pad - cold water circulating pad, one unit per package. The pads are used to provide a patient with local cold therapy by circulating chilled water through the pad that has been positioned on the patient.

Company
Cincinnati Sub-Zero Products Inc
Recall Initiated
January 7, 2015
Posted
February 19, 2015
Terminated
June 19, 2015
Recall Number
Z-1129-2015
Quantity
1390
Firm Location
Cincinnati, OH

Reason for Recall

Seals in the water path of the CT-99 cold therapy pad may separate during use, which could lead to a water leak.

Distribution

Nationwide Distribution including FL, IL, TN, GA, VA, MO, NM, WI, MI, NC, IN, OH, MA, CA, WA, PA, KY, MD, HI, and AL.

Lot / Code Info

Model: Catalog #CT -99 Part number: 50137 Lot number(s): 13380, 13390 and 13418.

Root Cause

Process design

Action Taken

Cincinnati Sub-Zero sent an Urgent Medical Device Recall Field Action Notice dated January 7, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately cease use and/or further distribution of the affected product. Customers were instructed to obtain a Return Material Authorization Number and return the pad(s) for exchange. Customers with questions were instructed to call 1-800-989-7373 or 513-772-8810. Customers were also asked to commplete and return the enclosed Response Form as soon as possible by fax to (513)772-9119 to acknowledge receipt of the notification and to inform CSZ that they have performed the requested actions. For questions regarding this recall call 513-772-8810, ext 6816