ES0120, AEM Disposable Low Profile Scissors, 3/4", 35cm, Sterile R, Encision Inc., Boulder, CO 80301 Cutting tissue during electrosurgical procedures

This recall has been terminated (originally issued May 31, 2011).

Company
Encision, Inc.
Recall Initiated
April 7, 2011
Posted
May 31, 2011
Terminated
August 9, 2011
Recall Number
Z-2398-2011
Quantity
233 Ten Count Boxes, 1 Kit
Firm Location
Boulder, CO
Official Source
View on FDA website ↗

Reason for Recall

Sterility may be compromised due to holes in packaging.

Distribution

Worldwide Distribution - USA (nationwide) and the countries of Australia and New Zealand

Lot / Code Info

Lot # QJQ, QKE, QKG, QLD, RAH, RBE

Root Cause

Package design/selection

Action Taken

Encision, Inc. sent an Important Safety Notice letter dated April 7, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to either carefully inspect any remaining pouches for readily visible holes in the Tyvek pouch, or return the affected lots to Encision for inspection. Customers were also asked to fill in and fax back the attached Confirmation to 303-444-2693.

More recalls by Encision, Inc.

View all recalls by this company →