Encision AEM Suction Irrigation Electrode, J-Hook, REF ES3884. Encision Inc., Boulder, CO 80301. The Suction Irrigation Electrodes and Adapters combine AEM shielding technology with suction and irrigation of fluids during laparoscopic electrosurgery to cut and coagulate tissue. The electrodes connect to a compatible electrosurgical generator via an adapter on the Encision AEM Monitor.

This recall has been terminated (originally issued August 30, 2011).

Company
Encision, Inc.
Recall Initiated
August 24, 2011
Posted
August 30, 2011
Terminated
September 4, 2012
Recall Number
Z-3096-2011
Quantity
83 units
Firm Location
Boulder, CO
Official Source
View on FDA website ↗

Reason for Recall

Electrode tips have been reported breaking off during use or cleaning.

Distribution

Worldwide Distribution: Nationwide distribution, and the country of New Zealand.

Lot / Code Info

Catalog: REF ES3884. Lots: QI, RA, and RE

Root Cause

Device Design

Action Taken

Encision Inc. sent out an "Urgent Product Recall" letter dated August 24, 2011 to all affected customers. The letter included: affected product, problem and instructed customers to discontinue use and return all affected electrodes to Encision. An enclosed reply form was to be faxed back to Encision indicating status of firm's inventory. For information on this recall call Encision at (303) 339-6917.