Accu-Chek Advantage glucose test strips; part numbers 336, 556, 787 and 966.

This recall has been terminated (originally issued April 9, 2003).

Company
Roche Diagnostics Corp.
Recall Initiated
February 12, 2003
Posted
April 9, 2003
Terminated
September 15, 2003
Recall Number
Z-0716-03
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

Crack in the bottom of the test strip vial will cause erroneously low or high blood glucose readings.

Distribution

United States, Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Denmark, Finland, France, Germany, Guatemala, Honduras, Hong Kong, Indonesia, Japan, Malaysia, Mexico, New Zealand, Norway, Pakistan, Peru, Philippines, Portugal, Singapore, South Africa, South Korea, Spain, Taiwan, Thailand, United Kingdom and Vietnam.

Lot / Code Info

All lots.

Root Cause

Other

Action Taken

Urgent Product Correction letters dated 2/12/03 were issued to pharmacists, specific health care providers, and the laboratory manager at customer accounts. Letter recipients were asked to advise their patients to inspect their vial for a crack prior to each use.