Siemens Mevatron Linear Accelerator Systems: KD2, K 7767, K 8067, KD, KDS and KDS2, Mevatron KD2, Part No.: 1940753; Mevatron KD2, Part No.: 08515520; Mevatron KD2-Part No.: 9822685; Mevatron K 7767, Part No.: 5694302; Mevatron K 8067, Part No.: 5694401; Mevatron KD, Part No.: 8319857; Mevatron KDS, Part No.: 9401522; Mevatron KDS2, Part No.: 9411588; and Mevatron KDS2, Part No.: 9822693, Siemens Medical Solutions USA, Inc., Concord, CA
- Recall Initiated
- July 27, 2007
- Posted
- February 19, 2008
- Terminated
- April 29, 2011
- Recall Number
- Z-0851-2008
- Quantity
- 1,284 total for recalls Z-0850-0852-2008
- Firm Location
- Concord, CA
Reason for Recall
Door Hinge failures; loose or falling off (Stationary Structure Doors)
Distribution
Worldwide- USA including the states of Missouri, Georgia, Delaware, Texas, Ohio, Alabama, Kentucky, Florida, California, North Carolina, New Jersey, New York, Illinois, Pennsylvania, Colorado, Tennessee, Arizona, Pennsylvania, Virginia, New Jersey, Michigan, Wisconsin, Indiana, Nebraska, Montana, Kansas, Mississippi, Rhode Island, Washington, South Carolina, Maryland, New Hampshire, Maine, Arkansas, Idaho, Missouri, Louisiana, Oklahoma, North Dakota, Minnesota, Massachusetts, and The District of Columbia and countries of Canada and Mexico.
Lot / Code Info
All Serial Numbers
Root Cause
Component design/selection
Action Taken
Stationary Door Hinges will be replaced on each device by the recalling firm's technician visits.