RecallDepth

MagNA Pure LC Instrument; Catalog number 2236931.

This recall has been terminated (originally issued July 23, 2003).

Company
Roche Diagnostics Corp.
Recall Initiated
June 26, 2003
Posted
July 23, 2003
Terminated
November 19, 2003
Recall Number
Z-1044-03
Quantity
17
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

A software bug in version 3.0 may result in prolonged mixing time, evaporation of elution volume and biased sample results for various protocols.

Distribution

Nationwide and Canada California, Georgia, Illinois, Maine, Michigan, Mississippi, New York, Ohio, Oregon, Pennsylvania, Texas, Utah, Wisconsin and Canada.

Lot / Code Info

All systems using software version 3.0

Root Cause

Other

Action Taken

An urgent product correction letter dated 6/26/03 was sent to each customer advising them not to use DNA protocols until the software update has been installed.

More recalls by Roche Diagnostics Corp.

View all recalls by this company →