RecallDepth

4Z1c Matrix Array Transducer, model number 10033682, used with ACUSON SC2000 ultrasound systems, model number 10433816, manufactured by Siemens Medical Solutions, Ultrasound Business Unit, Mountain View, CA. The 4Z1c transducer is a special purpose ultrasound transducer designed to help clinicians perform adult echocardiography and pre-operative and intra-operative surgical assessments. Secondary applications include pediatric echocardiography, primarily benefiting older and larger pediatric populations, general abdominal and intra-operative pediatric abdominal imaging and peripheral vessel imaging.

This recall has been terminated (originally issued May 5, 2011).

Company
Siemens Medical Solutions Usa, Inc.
Recall Initiated
July 10, 2009
Posted
May 5, 2011
Terminated
May 5, 2011
Recall Number
Z-2149-2011
Quantity
42 units
Firm Location
Mountain View, CA
Official Source
View on FDA website ↗

Reason for Recall

Burn risk-- due to coolant leaks, patients may suffer burns or blisters due to increased temperature.

Distribution

Nationwide. Foreign: Japan, Italy, South Korea, Singapore, Yemen, Germany, Australia, and Greece.

Lot / Code Info

seial numbers

Root Cause

Equipment maintenance

Action Taken

A customer safety advisory notification was sent out to customers on 7/10/09, and a mandatory software update will also be conducted.

More recalls by Siemens Medical Solutions Usa, Inc.

View all recalls by this company →