Pharmacy compounding system . It is a twenty-four source compounder for multi-source fluid mixing of both macro and micro ingredient down to 0.2 ml quantities. Product Usage: Pharmacy compounding System for parenteral macro and micro ingredients

This recall has been terminated (originally issued April 9, 2012).

Company
Baxa Corporation
Recall Initiated
March 23, 2009
Posted
April 9, 2012
Terminated
April 10, 2012
Recall Number
Z-1389-2012
Quantity
1322 units
Firm Location
Englewood, CO
Official Source
View on FDA website ↗

Reason for Recall

Software issue with compounding pump resulting in over delivery of component ingredients during routine operations.

Distribution

Worldwide Distribution - USA Nationwide including Canada.

Lot / Code Info

2 model numbers 2400 D and 2400 M.

Root Cause

Software design

Action Taken

Baxa sent a Safety Alert letter dated March 23, 2009 to all affected consignees. The letter identified the affected product, problem and recommended actions with specific instructions related to the issue. Additionally, the firm will be providing a warning label for the affected product as soon as possible and is currently working on a product upgrade to correct the issue. For assistance contact Baxa Technical Support at 800-678-2292. The firm is also providing a 24 hour a day, 7 days a week Technical Support for the affected product.