Access Hybritech p2PSA reagent, REF B03704 and REF A49752.

This recall is currently active, issued June 30, 2023. It was issued by Beckman Coulter, Inc..

Company
Beckman Coulter, Inc.
Recall Initiated
May 17, 2023
Posted
June 30, 2023
Recall Number
Z-2043-2023
Quantity
987 kits
Firm Location
Chaska, MN
Official Source
View on FDA website ↗

Reason for Recall

Reagent lot numbers include an insufficient concentration of blocking reagent.

Distribution

Distribution was made to CA, CO, CT, FL, GA, IL, IN, MA, MI, NC, NE, NJ, NY, OH, SC, TN, TX, UT, and WA. There was government distribution but no military distribution. Foreign distribution was made to Australia, Austria, Belgium, Brazil, China, Czech Republic, France, Germany, Hong Kong, India, Italy, Japan, Korea, Lithuania, Macao, Mexico, Portugal, Russia, Serbia, Singapore, Slovakia, Spain, Switzerland, Taiwan, and Vietnam.

Lot / Code Info

Lot #234320, exp. 7/31/2023, UDI (01)15099590211325(17)230731(11)220731(10)234320; Lot #234134, exp. 6/30/2023, UDI (01)15099590211325(17)230630(11)220630(10)234134; and Lot #234133, exp. 6/30/2023, UDI (01)15099590204112(17)230630(11)220630(10)234133.

Root Cause

Nonconforming Material/Component

Action Taken

The recalling firm issued letters to U.S. consignees dated 5/15/2023 via first class mail and email on 5/17/2023. The letter described the issue, impact, and the action to be taken. The consignee was instructed to discontinue using the reagent lots that were listed and discard all remaining reagent packs from the lots. The letter was to be shared with the consignee laboratory and/or medical director regarding the need to review previous patient test results. The Beckman Coulter representative was to be contacted for replacement product requests. If the product was forwarded to another laboratory, the consignee is to provide them a copy of the letter. The consignee is requested to respond within 10 days of receipt of the letter either electronically or manually by completing the enclosed Response Form. The letter for the OUS consignees was dated 5/16/2023 and was similar to the letter for the U.S. consignees.