Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry Analyzer

Company
Beckman Coulter, Inc.
Recall Initiated
October 1, 2025
Posted
November 4, 2025
Recall Number
Z-0428-2026
Quantity
80 units
Firm Location
Chaska, MN

Reason for Recall

An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.

Distribution

Worldwide

Lot / Code Info

A71460, UDI/DI 15099590369224, Serial Numbers: 902515, 902529, 902528, 902534, 902526, 902527, 902523, 902522, 902524, 902505, 902578, 902579, 902545, 902530, 902531, 902581, 902504, 902514, 902525, 902535, 902566, 902516, 902518, 902532, 902555, 902558, 902564, 902519, 902521, 902559, 902511, 902549, 902554, 902547, 902582, 902583; A30260, UDI/DI 15099590341602, Serial Numbers: 902536, 902541, 902517, 902563, 902565, 902557, 902513, 902507, 902543, 902560, 902562, 902506, 902508, 902574, 902533, 902572, 902540, 902569, 902568, 902577, 902510, 902580, 902567, 902573, 902570, 902571, 902538, 902553, 902552, 902512, 902539, 902561, 902575, 902544, 902542, 902556, 902537, 902509, 902520, 902546, 902548, 902551, 902550, 902576.

Root Cause

Process control

Action Taken

Beckman Coulter issue an URGENT MEDICAL DEVICE RECALL notice to its consignees on 10/01/2025 via USPS and email. The notice explained the issue, potential impact, and requested the following: If Wash PnP and/or Incubator PnP Z-motion errors appear in the analyzer event log, take the following steps: 1. Attempt to reinitialize the analyzer: a. Go to the Diagnostics screen on the User Interface. b. Select Initialize System. 2. If reinitialization is unsuccessful, contact Beckman Coulter Technical Support in your region. Beckman Coulter recommends posting this letter on or near the affected analyzers. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected products listed above to another laboratory, please provide them a copy of this letter. A field service representative will contact your laboratory to schedule a time to replace the affected parts. If you have any questions regarding this notice, please contact our Customer Support Center website: http://www.beckmancoulter.com